M700 Specifications

M700 (20K)StimulatorScreen

Part hemispherical bowl, radius 30cm integrated diffusing surface.

Test Fields

Central 30º : 100 points
Full 50° : 164 points
Peripheral 30°-50° : 73 points
Macula 10° : 49 points
Glaucoma 22°/50° : 104 points
Neurological 50° : 164 points
Quickscan 22°/30° : 40 points
Driving 50°/80° : 106 points
Binocular 30°/40° : 21 to 128 points
Binocular Driving Test 160° : 124 points
Spatially Adaptive Test 50° : 40 points

Stimulus Source

Read projection light emitting diode.

Stimulus Colour

Pale Green - wavelength 565nm
half bandwidth 28nm

Stimulus Size

Goldmann Size 111 (0.43°)

Stimulus Intensity

0.03 asb to 1000 asb
in 15 x 3dB steps / 45 x 1dB steps

Stimulus Duration

Adjustable 0.1 to 9.9 sec (nom. 0.2 sec)

Patient Response Time

(a) Adaptive to patient
(b) Operator selection of normal or slow ranges
(c) Adjustable 0.1 to 9.9 sec (nom. 1.1 sec)

Minimum Inter-stimulus Delay

Adjustable 0.1 to 9.9 sec (nom. 0.4 sec)

Background Illumination

10 asb (3.2cd/m²), automatic level control

Test Lens Diameter

38mm

Fixation Method

"Heijl-Krakau" blind-spot method, automatic tracking during test with visualand audible warning of fixation errors. Optional video camera.

Stimulator Unit Dimensions

626mm wide x 438mm deep x 713mm high

Stimulator Unit Weight

16kg approx

Stimulator Unit Power

110/220/240 VAC
50/60Hz 30W

Recommended System Requirements


N2100 TickN2100   CE